ACQUITY UPLC/UV Qualification Standards Kit [700002642]

  • Product
  • Documents
  • Columns & Vials
  • Systems

ACQUITY UPLC/UV Qualification Standards Kit [700002642]

The ACQUITY UPLC/UV Qualification Standards Kit is intended for performing system-level performance qualification on an ACQUITY System with TUV or PDA detectors, with or without an ELS detector in line, using the standard 10-mm flow cell configuration. This kit is designed for use by a Waters certified engineer or user as part of a performance maintenance or qualification activity. Protocols are available via the purchase of Waters qualification services or self directed qualification training and license.

The kit contains standards appropriate for verification of system reproducibility, flow rate, gradient performance, detector linearity, detector accuracy, injector accuracy, injector carryover.

Kit Contents:

  • Caffeine at the following concentrations: 0.01 mg/mL, 0.03 mg/mL, 0.04 mg/mL, 0.05 mg/mL, 0.08 mg/mL, 0.1 mg/mL, 0.16 mg/mL, 0.20 ug/mL and 4.0 mg/mL in 90/10 water/acetonitrile, 10 mL per ampule
  • One ampule with 0.03 mg/mL thiourea, 0.110 mg/mL acetophenone in 90/10 water/acetonitrile, 10 mL
  • One ampule with 4.0 μg/mL each of 2-acetylfuran, acetanilide, acetophenone, propiophenone, butylparaben, benzophenone and valerophenone in 90/10 water/acetonitrile, 10 mL
  • Two 10 mL ampules of 90/10 water/acetonitrile 

 

Store at room temperature.

  • ACQUITY UPLC BEH C18 Method Validation Kit, 130Å, 1.7 µm, 2.1 mm X 50 mm, 3/pkg (186004044)

System Performance Standards Key

Set Up

Setup standards are used upon installation of a new system, start-up of a system which has been idle for a period of time, or to re-start a system that has been moved to a new location. These standards may include multiple components within a kit which allow step by step setup and calibration, or they can be single mixtures which will enable a full system assessment and initialization.

Calibration

These standards can be used for instrument calibration. Calibration is usually performed at a frequency that minimizes the risk of being out of calibration, in addition to major events suc h as when the instrument is moved and upon any major repair. Calibration can be confirmed as often as your standard operating procedures recommend. Look for solutions to calibrate your MS mass accuracy and resolution, UV detector linearity, UV and Fluorescence detector wavelength accuracy, injector accuracy and precision, and solvent system flow rate accuracy.

Performance

Your system is comprised of many interdependent components working together. An issue with any one component can produce erroneous final results. All components performing correctly will produce results within an expected variability. Some of the system components include mobile phase, column, tubing, pumping, injecting, temperature controlling, detecting, data collection rate, and integration
parameters. An issue in any one of the components listed can affect your result.
 
Performance Check Standards should be utilized on a routine basis. This includes anytime a column is changed, a method is changed, the instrument is moved, or confirmation of system performance is desired. Confirming the performance of your system can be done on a system basis with suitability standards or per individual instrument component. Benchmark performance should be established after
calibration and confirmed regularly.

Tuning

These Mass Spectrometry solutions are used routinely and are available for every ionization mode. Use the solution to optimize the mass spectrometer setting for optimal signal either manually or through Intellistart.

Hazardous

This product contains hazardous material which requires special freight handling. Additional charges may apply.

Have a Question?

Get Expert Advice!

Contact Us | Sales Offices

Related Product

ACQUITY UPLC/UV Start Up Standards Kit [700002669]

The ACQUITY UPLC/UV Start Up Standards Kit is intended to perform system precision and gradien ...