GCT Premier(EI,CI,DCI,SIP) Standards Kit [700002878]

  • Product
  • Documents
  • Columns & Vials
  • Systems

GCT Premier(EI,CI,DCI,SIP) Standards Kit [700002878]

The GCT Premier(EI,CI,DCI,SIP) Standards Kit is intended to assess the performance of the GCT Premier. This kit is designed for use by a Waters certified engineer or user as part of a performance maintenance or qualification activity.  Protocols in the use of this kit are available via the purchase of Waters qualification services or self directed qualification training and license.  

The kit contains 12 Components in the following formats:

  • 2 ampules of 1 ml Heptacosafluorotributylamine for Resolution Specification
  • 2 ampules of 1 ml 2,4,6 - Tris (trifluoromethyl) - 1,3,5 - triazine for Tuning and Calibration  
  • 2 ampules of 1 ml Chloropentafluorobezene for Tuning
  • 2 ampules of 1 ml of 10 ng/μL Methyl Stearate in Hexane for GC-MS Mass Measurement Accuracy
  • 2 ampules of 1 ml of 50 ng/μL Methyl Stearate in Hexane for DIP EI +ve Mass Measurement Accuracy
  • 2 ampules of 1 ml of 1 pg/μL Hexachlorobenzene in Hexane for GC-EI +ve Sensitivity Specification
  • 2 ampules of 1 ml of 2.5 ng/μL Hexachlorobenzene in Hexane for DRE demonstration
  • 2 ampules of 1 ml of 100 pg/μL of Benzophenone in Hexane for GC-CI +ve Sensitivity Specification
  • 2 ampules of 1 ml of 200 ng/μL Benzophenone in Hexane for DIP CI +ve Mass Measurement Accuracy
  • 2 ampules of 1 ml of 1 pg/μL Octafluoronaphthalene in Hexane for GC-CI –ve Sensitivity Specification
  • 2 ampules of 1 ml of 50 ng/μL of Octafluoronaphthalene in Hexane for DIP CI -ve Mass Measurement Accuracy
  • 2 ampules of 1 ml of 20 ng/μL Caffeine in Hexane for DCI +ve Mass Measurement Accuracy
  • 20 x 32 mm glass screw top vials
  • 20 x 5.75” glass Pasteur pipettes
  • 2 mm latex pipette bulb

 

Store at room temperature.

System Performance Standards Key

Set Up

Setup standards are used upon installation of a new system, start-up of a system which has been idle for a period of time, or to re-start a system that has been moved to a new location. These standards may include multiple components within a kit which allow step by step setup and calibration, or they can be single mixtures which will enable a full system assessment and initialization.

Calibration

These standards can be used for instrument calibration. Calibration is usually performed at a frequency that minimizes the risk of being out of calibration, in addition to major events suc h as when the instrument is moved and upon any major repair. Calibration can be confirmed as often as your standard operating procedures recommend. Look for solutions to calibrate your MS mass accuracy and resolution, UV detector linearity, UV and Fluorescence detector wavelength accuracy, injector accuracy and precision, and solvent system flow rate accuracy.

Performance

Your system is comprised of many interdependent components working together. An issue with any one component can produce erroneous final results. All components performing correctly will produce results within an expected variability. Some of the system components include mobile phase, column, tubing, pumping, injecting, temperature controlling, detecting, data collection rate, and integration
parameters. An issue in any one of the components listed can affect your result.
 
Performance Check Standards should be utilized on a routine basis. This includes anytime a column is changed, a method is changed, the instrument is moved, or confirmation of system performance is desired. Confirming the performance of your system can be done on a system basis with suitability standards or per individual instrument component. Benchmark performance should be established after
calibration and confirmed regularly.

Tuning

These Mass Spectrometry solutions are used routinely and are available for every ionization mode. Use the solution to optimize the mass spectrometer setting for optimal signal either manually or through Intellistart.

Hazardous

This product contains hazardous material which requires special freight handling. Additional charges may apply.

Have a Question?

Get Expert Advice!

Contact Us | Sales Offices

Related Product

Resolution Check on all GC-MS products [186006862]

The Resolution Check on all GC-MS products is intended to check the resolution and for routine ...

Related Product

CI optimization on GCTs only [186006863]

The CI optimization on GCTs only is to be used to optimize the chemical ionization signal and ...

Related Product

CI/FI optimization and Tuning on GCTs only [186006864]

The CI/FI optimization and Tuning kit on GCTs only is intended to optimize the chemical ioniza ...

Related Product

FI optimization on GCTs only [186006865]

The FI optimization on GCTs only standard is intended to optimize the chemical ionization sign ...